Tokai Hit stage top incubator

Last modified by mliljest@helsinki_fi on 2024/01/24 07:25

Incubation system with STXG controller unit and WSKMX stage top chamber. Temperature, humidity and CO2 control.

Suggested amount of dH2O in water bath: 32ml

Temperature set points:

S/N

35mm dish

Well plate

Top

43.0 °C

43.5 °C

Bath

41.0 °C

41.0 °C

Stage

38.0 °C

37.5 °C


Lens heater set values according to objective (WD, Imm) and dish bottom material:

Magnification

Objective

NA

Imm.

Glass bottom

Plastic bottom

5x

N Plan

0.12

Dry

OFF

OFF

20x

HC PL Fluotar Corr

0.4

Dry

OFF

OFF

20x

HC PL Apo

0.8

Dry

36 °C

39 °C

40x

HC PL Fluotar

0.8

Dry

36 °C

39 °C

40x

HC PL Apo CS2

1.3

Oil

41 °C

-


All set values from instrument manual and have not been confirmed by BIU. Use your own discretion.