Leica Stellaris 8 FALCON
Intruduction
Please find the introduction video on our YouTube channel.
https://www.youtube.com/watch?v=7BLrn7Cuicg
Technical details:
MICROSCOPE:
Leica DMI8
Fully motorized inverted microscope.
SuperZ-galvanometric stage
motorized xy-stage
AFC hardware autofocus
OKO lab Cage incubation system including CO2 Unit with active humidity controller.
OBJECTIVES:
HC PL APO 10x/0.40 CS2 (air, wd 2.56mm)
HC PL APO 20x/0.75 CS2 (Air, wd 0.62mm)
HC PL APO 63x/1.20 W motCORR CS2 (water, wd 0.30 mm)
HC PL APO 40x/1.25 motCORR (glycerol, wd 0.35mm)
The MotCORR objectives have a motorized correction collar for coverslip thickness correction
Motorised DIC contrast for 63x water immersion objective (IC Prism E, IC Condencer Prism K10)
LASERS:
White Light Laser Stellaris 8 (440-790nm)
UV (diode 405 nm/50mW)
Acusto Optical Beam Splitter AOBS
DETECTION:
Fluorescence
Two HyD S (positions 1 and 3) most sensitive detector at 410 - 560 nm range. Tolerant to over exposure and may be used for coverslip correction and other reflected light imaging.
Two HyD X (positions 2 and 4) most sensitive detector at 560 - 740 nm range. Not tolerant to over exposure, shuts down and may be permanently damaged. Fast detector optimized for FLIM, also has better image quality with resonant scanner compared to HyD S.
HyD R (position 5) most sensitive detector above 740 nm. Not tolerant to over exposure, shuts down and may be permanently damaged. Fast detector optimized for FLIM, also has better image quality with resonant scanner compared to HyD S.
Background information on the detectors
Transmitted light
1 PMT
FOV scanner and Resonant 8kHz Tandem Scanner
Information on Fluorescence Lifetime Imaging (FLIM)
FALCON Fast Fluorescent Lifetime measurement
Fluorescence lifetimes for Alexa Fluor Dyes
http://www.iss.com/resources/reference/data_tables/LifetimeDataFluorophores.html
Software LAS X:
4.0.2. (18.9.2020)
4.1.1. (17.11.2020)
4.3.0 (6.8.2021)
4.4.0 (22.11.21)
4.5.0 (18.7.22)
4.6.0 (22.8.23)
4.7.0 15.1.2024
Quick start manual
DSE manual
Dynamic signal enhancement is meant to allow fast image acquisition with passable image quality.
Tau-gating manual
Theoretical background and practical instructions on how to use lifetime based detection gating to reduce background in your images.
TRAININGS
As Leica Stellaris 8 has more functionalities than your average confocal microscope, a regular three-hour training is not enough to learn all the ins and outs of the system. For this reason, the trainings are divided into four modules and the user can chose which module(s) they are trained on, each module taking two to three hours depending on the content. It is possible to be trained for more than one module, but each such training will be organized as a separate instance.
Modules and their contents are listed below
Quantification module
-Basic use
-Photon counting / gating
-Navigator
FLIM/FRET module
-Basic use
-FLIM/FRET measurements
-Handling and manipulating FLIM/FRET data
-Spectral separation based on lifetimes
Lightning module
-Basic use
-Lightning
Live-imaging module
-Basic use
-Resonance scanner
-Navigator