Leica SP8 X
Last modified by efazeli@helsinki_fi on 2024/01/24 07:25
Known issues and possible solutions:
Sample not visible through oculars
Symptom: No fluorescent nor bright field light via oculars.
Cause: When switching from detectors to oculars, the light path does not turn correctly every time.
Solution: On the touch panel, (1) select the Objective tab and (2) switch Port to "Eyepieces 100 %"
Frequency: Has happened about 5 times (by 23.8.2019). Seems to be happening more and more often.
SOLVED: pieces of coverslip glass and other dirt cleaned from inside the microscope!