FRAP How-to on Leica Confocal Microscopes
Last modified by Jussi Kenkkilä on 2024/11/04 14:41
For TCS SP8 CARS, TCS SP8 X and Stellaris 8 FALCON/DLS
Short Instructions
- Open FRAP Wizard from the TCS SP8/Stellaris 8 tab
- Follow the steps in Overview tab
- Setup imaging parameters for pre- and post-bleach imaging
- Only use PMT or HyD S detectors
- NOT HyD, HyD X or HyD R
- The bleaching laser intensity can damage other detectors
- Choose Acquisition Mode (xyt or xyzt)
- Set Acquisition Speed (fast speed is preferred)
- Select laser lines for photobleaching
- On Stellaris you can use the Argon 488 nm line for more power (Drag the laser excitation to exactly 488 nm to choose it instead of the white light laser)
- You can select several lines if needed
- Use 80-100% laser power
- Turn on the FRAP Booster for more intensity
- Use Fly Mode to for better time resolution
- Zoom In to the FRAP area for better photobleaching if not in Fly Mode
- Set Background to Zero if using Fly Mode
- Delete Bleach Images After Scan (if you’re not using the Fly Mode)
- When you want to use different laser lines for several ROIs, uncheck Use Laser Settings For All ROIs and open ROI Configuration to choose the lasers
- Define the number of pre-bleach, bleach and post-bleach intervals in the TimeCourse tab
- You can have several post-bleach steps with different number of frames and frame intervals
- Minimize for best time resolution
- Match the acquisition speed to expected recovery time; you should have at least 10 datapoints by half recovery
- Run the time series as long as there is increase in fluorescence intensity
- Save your settings if you want to repeat the same experiment