Rotronic radiation shield & MP102H modifications
The radiation shield modifications:
-Junction box removed
-The hole from the arm to the junction box widened close to the inner width of the arm
-The connector for the fan power removed and an extension piece with a 5-pin M12 male connector reaching about 10 cm out from the arm added (+24 V connected to pin 1 and GND to pin 2)
-A power cord of the required length with a 5-pin M12 female connector with the same pin configuration added
The MP102H modifications:
-Shield connected to the power ground inside the sensor
-The cord cut to the reach about 10 cm out from the arm
-A Binder 712 male connector added with the connections listed below
-A Sabix D 315 FRNC 12*0,14 mm2 cable of the required length with a Binder 712 female connector added with the connections listed below
MP102H-530300 | Binder 712 | Sabix | |
Color | Wire | Pin number | Color |
Gray | Power GND | 1 | Gray |
Green | Power + (5-24 V) | 2 | Green |
White 1 | Out1 + (RH) | 3 | White |
White 2 | Out1 - (A-GND1) | 4 | Gray-pink |
Brown | Out2 + (Temp.) | 5 | Brown |
Pink | Out2 - (A-GND2) | 6 | Pink |
Red | RS485 (TxD) | 7 | Red |
Blue | RS485 (RxD) | 8 | Blue |
Black | PT100 PT-S | 9 | Black |
Purple | PT100 PT-AS | 10 | Purple |
Orange | PT100 PT-R | 11 | Red-blue |
Yellow | PT100 PT-AR | 12 | Yellow |
Shield | - | - |
|
Table 1. MP102H wire connections
New sensors have been measured for comparison with Ekis PT100 with a Delta-T DL2e logger and a 4-wire card with the wire configuration listed in table 2. In one channel there have also been used a VFR S102CT 100 ohm (±0,01%) precision resistor to track possible errors.
Wire | Color | Pin |
PT-S | Black | - |
PT-R | Red-blue | + |
PT-AS | Puple | GND |
PT-AR | Yellow | I |
Table 2. Wire connections for measurement