Rotronic radiation shield & MP102H modifications

Last modified by tematil@helsinki_fi on 2024/02/07 06:31

The radiation shield modifications:

-Junction box removed

-The hole from the arm to the junction box widened close to the inner width of the arm

-The connector for the fan power removed and an extension piece with a 5-pin M12 male connector reaching about 10 cm out from the arm added (+24 V connected to pin 1 and GND to pin 2)

-A power cord of the required length with a 5-pin M12 female connector with the same pin configuration added

 

The MP102H modifications:

-Shield connected to the power ground inside the sensor

-The cord cut to the reach about 10 cm out from the arm

-A Binder 712 male connector added with the connections listed below

-A Sabix D 315 FRNC 12*0,14 mm2 cable of the required length with a Binder 712 female connector added with the connections listed below

 

MP102H-530300

Binder 712

Sabix

Color

Wire

Pin number

Color

Gray

Power GND

1

Gray

Green

Power + (5-24 V)

2

Green

White 1

Out1 + (RH)

3

White

White 2

Out1 - (A-GND1)

4

Gray-pink

Brown

Out2 + (Temp.)

5

Brown

Pink

Out2 - (A-GND2)

6

Pink

Red

RS485 (TxD)

7

Red

Blue

RS485 (RxD)

8

Blue

Black

PT100 PT-S

9

Black

Purple

PT100 PT-AS

10

Purple

Orange

PT100 PT-R

11

Red-blue

Yellow

PT100 PT-AR

12

Yellow

Shield

-

-

 

Table 1. MP102H wire connections

 

New sensors have been measured for comparison with Ekis PT100 with a Delta-T DL2e logger and a 4-wire card with the wire configuration listed in table 2. In one channel there have also been used a VFR S102CT 100 ohm (±0,01%) precision resistor to track possible errors.

 

Wire

Color

Pin

PT-S

Black

-

PT-R

Red-blue

+

PT-AS

Puple

GND

PT-AR

Yellow

I

Table 2. Wire connections for measurement