AMS logbook spring 2008

Last modified by junninen@helsinki_fi on 2024/02/07 06:30

----------------------------------------------------------------------------------
 29.Apr.2007
 20:00 start sampling
 Chopperi is not working => no EI, no bitwise, no ptof
 He 7%
 threshold 4runs start at 1

20:30 previous runs might not be reliable, sampling line fan was off
 20:40 filter #9
 21:10 sample #15

----------------------------------------------------------------------------------
 30.Apr.2007
 7:50 stoped to put UPS
 11:56 back up
 oven 600C from 540C
 12:13 autosave data on,#143

3.May.2008
 09:40 Heater temp 540 and decreasing all the time. Problems has probably
 started couple of hours ago. All the flows are ok.
 Filament1 was broken (no light in the circle window). After Heikki's telephone consultation
 filament2 is been used.
 Air conditioner is set to cool the room to 22 degrees. Before it was heating the room to
 27 degrees. HM
 12:27 Heater temp 573.

5.5
 18:07 Shut down to replace pump nr 6.
     Replaced pump and checked that the chopper wasn't really spinning when in "bad mode".
     Letting pump down over night.

6.5
 9:00     Everything looked ok, except that when doing m/z calibration, no difference between open and closed.
     Also the sound of changing open->closed had changed. So I opened it again and checked that the servo
     seemed to be working.
 9:40    Now pumping down again.
 9:53    Heater and filament on.
 11:50    Voltages back on. Suddenly realized that the servo hadn't been the problem, but just me being stupid and testing with the inlet closed!!!
 12:23    m/z calibration, and tuning ready. Back to measuring.
     He flow now at 4.5 mLPM.

7.5
     Seems like there was no data files produced after the data backup in the morning.
     Only after stop and start of filter measurements did system start saving again.

18:05    Run 1997, filter inline.
 19:17    Run 2011, filter removed. Back to ambient.

8.5

10:55    2196, ammonium sulfate generation.
     Medium DMA, sheath flow 10 LPM, V = 870V = 100 nm, CPC = 660 cm-3.
     2198, changed save time from 5min to 1 min. Also slightly smaller particles.
     2199, V=1250V = 110nm, CPC = 400
     2205-2227  Made stronger solution. V=6600V = 350nm. Conc fluctuating a lot, but is luckily saved.
     2231, Switch to AN., Still at 350 nm. Switching concentrations.
 12:58    2306, back to ambient. Saving time was still on 60s.
 13:37    2343, Saving time 180s
     

8.5

11:18    2772, MCP 2100->2200V

16.5

backup ME

19.5

10:30 backup ME

20.5

12:20 backup Me
 12:50  8021 , MCP voltage 2200 -> 2250

21.5

11:14 backup ME

20:57 filter #8655 140sec
     #8656 first full filter
 21:29    #8665 last full filter
 #8666 at 22 second sample back on
 #8668 last full sample
 21:39 qstop and try SI with tofdaq

SI = 9.8 mV*ns

22:26 sampling

==========================================
 22.05.2008

7:19 stop pumps for chopper repair
 sampling valve closed
 He flow 80ml/min
 chopper is on, display 24

10:56 chopper display 25
 open pump 4 valve, clearly under vacum
 chopper 24, copper on/off has no effect
 opened flange, chopper is not spinning
 have to deplug the cable to get chopper out
 immitiately when pluged back, chopper started to spin
 I FOUND THE PROBLEM!!!!!!!!!!!!
 motor wire connection very close to the motor!
 I'll change the motor

14:32 chopper working and pump down the system

18:?? ahh stupid toggle problam again, many waisted hours...
 21:33 tophat looks strange... still, resumed to sample with genalt mode, MS dwell 10 ptof dwell 10 savetime 300sec
 #8670
 =============
 23May08

tuning
 SI=9.8 (blocked signal mz40)
 MPC=2300

14:24 #8819 resume to sample

26.5

8:30 backup ME

27.5

08:35 backup JM
 10:42 backup ME kun ei huomannut ett oli jo tehty

28.5

8:00 backup ME

29.5

9:00 backup ME

30.5

9:36 backup JM
 13:30 backup ME

14:25 Set baseline, no real change
 SI 44/40:
 8.8/9.7 ; 7.8/8.6 ;7.5/8.3 ; 7.5/8.2
 MCP 2300->2375
 8.9/9.7 ; 8.8/9.6 ; 8.6/9.6 ; 8.7/9.6

2.6
 14:20 backup ME

3.6
 9:10 backup ME

4.6
 8:05 backup ME
 SI 44/40: (beam blocked)
 8.1/8.8 ; 8.0/8.7
 MCP 2375->2450
 8.9/9.9 ; 8.9/9.9 ; 8.8/9.8
 Set SI to 9.8 (from 9.6)

6.6
 14:55 backup ME

9.6
 10:30 Noticed that servo had probably broken. Also the MFC was not working, possibly due to power failure on Monday morning
     Backed up the data. Turned off He.
     Will tell someone to go and close the inlet valve. Since servo is stuck in open,
     it may be possible to calibrate before shutting down.
     Pump data had not been saved since 23.5

12.6
 final calibration didn't work out since both filaments are broken
 tofams software will not start, -> reboot

actually, after restart filament woke up. so, it is not broken
also the servo works

13:20 SI with tofdaq
 MCP 2450 beam open, sample valve closed

14.1 mV*ns 7.22 bit*ns

13:30 SI with tofams
 MCP 2450 beam open, sample valve closed
 SI 44/40 (620/1530; 1611/4006 ions)
 8.2/8.1; 8.0/8.2 8.1/8.2 (bit*ns)

SI = 8.1

14:00 servo stoped working again
 restart and servo works again, strange ???????

Velocity calibration
 run/particles/dma guidancs voltage/sheath/sample flow/diameter
 #13633/321/3.34/10/1.1/250
 #13634/276/4.3/10/1.1/300
 #13635/1480/0.78/10/100

BFSP #13638
 1730 particles / cm3 CPC reading
 225 particles / cm3 AMS reading
 DMA voltage/guidance 983/0.785
 DMA sheath/aerosol 10/1.1
 particle diameter 100nm
 hit rate 3%

150 particles collected
 back and send to helsinki

SI again, sample valeve open, chopper blocked
 44/40
 9.1/9.8; 8.7/9.4; 8.5/9.3
 2844/14537; 2132/11334; 2598/13442

SI=9.0

MS mode
 DMA 0.76V/982V/100nm/600cm-3
 from run 13651
 DMA 2.416V/3000V/200nm/130cm-3
 17:47 servo stoped again
 restart and works again

MCP 2450->2500
 BFSP #13663
 145 particles / cm3 CPC reading
 -1030 cm-3 AMS reading
 DMA voltage/guide 2.415/3015/
 DMA sheath/aerosol 10/1.1
 particle diameter 200nm
 hit rate 2%
 filtering treshold 4
 122 particles
 sampling valve was closed, damn

valve open
 BFSP #13664
 130 particles / cm3 CPC reading
 80 cm-3 AMS reading
 DMA voltage/guide 2.415/3015/
 DMA sheath/aerosol 10/1.1
 particle diameter 200nm
 hit rate 1%
 filtering treshold 4
 65 particles

#13668 filter
 #13673 las full filter run, well actually not, servo had died at some point