Environmental chamber (microtomography laboratory)

Last modified by henrikma@helsinki_fi on 2024/02/07 06:20

Configuration of sensors and dew point generator:

Installing the chamber (Quick guide)

Click to expand

1) Turn CT operation of with the attached key

2) Unscrew the goniometer holder and attach the chamber-specific holder/plate

3) Turn CT operation on

4) In datosx acquisition, choose CNC → Drive reference → All axes

5) Move rotating stage and detector according to the instructions in the attached file below

6) Place the chamber on top of the rotating stage so that the sample holder fits too the bottom slit

    Be careful! Don't damage the detector, beryllium window or other components!

7) Attach the sample holder if not done yet

8) Connect the plastic tube from the dew point generator and all necessary censors (see instructions in the configuration manual

Kammion fyysiset asetukset (in Finnish).odt

Codes and programs:

(Coming soon...)