...
- 2 PMT detectors (One with extra cooling for increased sensitivity in infrared)
- 1 GaAsP detector
Airyscan detector
Transmitted light detector (T-PMT)
Software and accessories for live cell and multi-well imaging
- Zeis ZEN 2
- Modules: Experiment designer, Tiles & Positions, FRAP and FRET
- Definite Focus hardware autofocus system
- Scanning stage for multipoint or mosaic imaging
- Large environmental chamber
- Incubator insert PM S1 (Temperature and CO2 control): Observation area 32 x 30 mm (oval) https://www.pecon.biz/products/product/?pid=5271%7CHeating_Insert_P_S
- 40x and 63x water immersion objectives are available upon request
Environmental Chamber Temperature Setting
The readout of the temperature probe differs slightly from the media temperature in the sample chamber. To heat the sample to +37 C, adjust the Incubator to:
- H Insert P +38.0 °C
- H Unit XL +38.2 °C
Test your own fluorophores on this optical system
...
- 2 high sensitivity hybrid detectors (HyD)
- 1 PMT detector
- 2 reflected light non-descanned PMT detectors with the possibility for filters
- 2 transmitted light non-descanned PMT detectors with the possibility for filters
Software and accessories for live cell and multi-well imaging
- Full-enclosure type temperature-controlled environmental chamber for live-cell imaging and bleaching FRAP applications.
- An inverted temperature-controlled slow-flow imaging chamber available on request
...